HP 6890 GC with Single FID and Pyrolyzer

HP 6890 GC System includes single Split/Splitless Injector, single Flame Ionization Detector, Double-Shot Pyrolyzer PY-2020D, Electronic Pressure Control on detector and injector, HPIB Communications installed on GC, manuals, and NINETY (90) DAY WARRANTY. Autosampler and Data System available.

Additional or alternative detectors available.

HP 5890/6890 GC Starter Kit


Specifications

Data communications

  • HP-IB, RS-232-C, and two analog output channels (1-mV, 1-V, and 10-V output available) as standard
  • INET interface optional

Heated zones

  • Independent heated zones, not including oven: six (two inlets, two detectors, and two auxiliary)
  • Maximum operating temperatures for auxiliary zones: 400° C

Column oven

  • Operating temperature: 4° C above ambient to 450° C
  • Temperature setpoint resolution: 1° C
  • Maximum run time: 999.99 min
  • Programming ramps/plateaus: 6/7

Inlets

  • All inlets include septum purge.
  • Split/splitless: electronic flow, 400° C max, electronic entry of pressure or flow and split ratio version.
  • Pressure setting range: 0-100 psi
  • Total flow setting range: split/splitless, 0-200 ml/min N2, 0-1000 ml/min H2 or He

Detectors: All detectors include electronic on/off of all detector gases.

    FID
  • 450° C maximum operating temperature
  • Automatic flame ignition from the keypad or ChemStation
  • Flame out detection
  • MDL: <5 pg carbon/s as propane using N2 carrier and 0.2794-mm jet
  • Linear dynamic range: < 10%, 107 with N2 carrier and 0.2794-mm jet
    NPD
  • 400° C maximum operating temperature
  • Automatic baseline adjusting via keypad or ChemStation
  • MDL: <0.4 pg N/s, <0.2 pg P/s with azobenzene/malathion mixture
  • Selectivity: 35000 to 1 gN/gC, 75000 to 1 gP/gC with azobenzene/malathion/ octadecane
  • Dynamic range: >105 N >105 P with azobenzene/malathion mixture
    TCD
  • 400° C maximum operating temperature
  • Single filament with microcell
  • MDL: <400 pg propane/ml He caxrier (MDL may be affected by laboratory environment.)
  • Linear dynamic range: 105 (5%)
    ECD
  • Equipped with anode purge for contamination resistance
  • 400° C maximum operating temperature
  • Makeup gas types: argon (5%)/methane, nitrogen
  • Radioactive source: 15 mCi 63Ni plated
  • MDL: <0.04 pg/s lindane
  • Dynamic range: > 104 with lindane

HP 5890/6890 GC Starter Kit

Includes RTX-5 column*, 50' 1/8" pre-cleaned copper tubing, tubing cutter, ten (10) .53 column ferrules, syringe, FID flow adapter, column cutter, column cutter, 25 septa, injector wrench, four (4) brass tees, ten (10) brass nuts and ferrules, 7/16" wrench, 1/2" wrench, 9/16" wrench and 1/4" wrench.

*RTX-5 (low-polarity phase; crossbond 5% diphenyl/95% dimethyl polysiloxane): .53 mm ID, 1.5 µm film thickness x 30 m length, -60 to 310/330° C temperature limits.