-科学分析仪器消耗品、配件服务

Professional Consumable&Parts Service Provider

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

欢迎光临北京思博全科技有限公司网站
GC&耗材&备件
仪器配件库
 
 
LC&耗材&备件-Waters HPLC
 
 

Waters LC Module I

This compact, fully integrated system is in excellent condition and tested at GenTech. Includes multisolvent delivery, automated sample injection, and high sensitivity UV/Vis detection for the high volume, high throughput laboratory. Fully automated and validated, the LC Module I is ideal for the regulatory-compliant laboratory that needs efficient and effective analyses. Preventive maintenance features, such as pump and seal wash, lamp save and self-diagnostic procedures, ensure the long-term integrity of the unit. The system features fast setup, unattended operation with automated start-up and shut-down routines, including priming and flushing and ability to program up to 96 vials of varying injection volumes and injections per vial. It incorporates a 715 Autosampler, 600 Pump & 486 Detector in one unit. Includes SIX (6) MONTH WARRANTY.

Pump Specifications
  • Number of solvents: One to four
  • Modes of operation: Gradient, isocratic, and flow programming
  • Flow range: 0.00 to 20.00 mL/min ( standard heads ), in 0.01 mL increments
  • Composition range: 0 to 100% programmable in 1% increments for each of four reservoirs ( A, B, C and D ). Total composition must sum to 100%
  • Compositional accuracy: Better than ±0.5%, independent of pressure
  • Automatic solvent sparging: Standard, helium gas, input gas pressure range 50 to 150 psi. Helium flow rate range 0 to 100 mL/min, Programmable to 1 mL/min
  • Pressure maxima:
    - 4500 psi at 10 mL/min
    - 1000 psi at 45 ml/min
  • Programmable pressure limits:
    - Upper 51 to 4500 psi
    - Lower 0 to 4450 psi
Autoinjector Specifications

  • Maximum pressure: 4500 psi
  • Sample capacity:
    - 48 vial carousel: 48 screw cap vials. Teflon self-sealing or Teflon disc septa
    - 96 vial carousel: 96 1- ml snap-in push cap
  • Minimum injection volume: 1.0 uL
  • Functions:
    - Replicate injections programmed from the controller
    - Autoprime
    - Discrete needle wash to prevent cross-contamination and also provide pump seal wetting
    - Programmable standard sequencing
  • Carryover: Less than 0.1% previous sample component concentration, except with 304 stainless steel parts
  • Cross contamination: Dependent on selection of needle wash solvent and sample interaction with 304 stainless steel parts
  • Reproducibility: Less than or equal to 1% for 1- to 100 ul injections
  • Linearity: Coefficient of determination greater than .99 from 1 to 100 uL

UV Detector Specifications
  • Wavelength range: 190 to 600 nm
  • Accuracy: ± 2 nm
  • Repeatability: ± 0.25 nm
  • Noise: 3.0 x 10 -5 AU (peak-to-peak, 1 second filter setting, dry cell, 254 nm)
  • Drift: 2.5 x 10 -4 AU with ambient temperature change of less than 1 °C/hour
  • Sensitivity setting range: 0.001 to 2.0 AUFS
  • Filter setting range:
    - Hamming: 0.1 to 5.0 seconds
    - RC: 0.1 to 99.0 seconds
  • Flowcell:
    - Type: Dual-beam Taper-Cell Design
    - Path length: 10 mm (standard, analytical)
    - Pressure limit: 1000 psi
    - Materials: 316 stainless steel Teflon, Suprasil II
  • Lamp source: Deuterium arc lamp
  • Lamp power supply: 309 mA Regulated Constant Current (fused at 3/8 A)
  • Lamp power: 30 W
Column Heater Specifications (If Equipped)
  • Temperature range: 0-99 °C
  • Humidity: 20 to 90% noncondensing

 

 


 
 
 
______________________________________________________________________________________